Model 2110 5½ Digit Multimeter Reference Manual
Section 5: Remote commands
2110-901-01 Rev. C/August 2013
5-19
Example
MEASure:VOLT[:DC]? 10,0.001
Measure 4 ½ digits on the 10 Vdc range.
MEAS:CURR:AC? 1,1E-5
Measure 5 ½ digits on the 1 A range.
Also see
(on page 5-16)
(on page 5-17)
:CONFigure:<function> [<range>|MIN|MAX|DEF,<resolution>|MIN|MAX|DEF]
(on page 5-14)
SCPI command subsystem reference
The following SCPI subsystems are available on the Model 2110. The subsystem commands are
summarized in the following tables.
•
CALCulate subsystem (Refer to
(on page 5-20))
•
DISPlay subsystem (Refer to
(on page 5-21))
•
SENSe subsystem (Refer to (
(on page 5-21))
•
STATus subsystem summary (Refer to
(on page 5-25))
•
SYSTem subsystem (Refer to
(on page 5-25))
•
TRIGger subsystem (Refer to
(on page 5-26))
This section includes commands that are device-specific to the Model 2110. Although not included in
the 1999.0 version of the SCPI standard, these commands are compliant to the SCPI format and they
follow the syntax rules of the standard. Device specific commands are indicated in the command
description.
Many of the required SCPI commands are accepted by the instrument, but are not described in this
manual for simplicity or clarity. Most of these non-documented commands duplicate the functionality
of a command already described in this section.