12
EN
Pulsair Intellipuff Tonometer by KEELER
5. Continue to move in slowly towards the
patient, two green dots will appear.
6. Continue to move closer, a red reflex
appears.
7. Move closer. At a distance of
approximately 15mm, a black cross on
red or ‘bow tie’ image appears. Centre
this image (on the central bar) and the
Pulsair IntelliPuff fires automatically.
8. Once you have taken a reading, remain in the operating position; wait a few seconds for
the air chamber to refill. When the ‘bow tie’ image appears the Pulsair IntelliPuff takes
a subsequent reading. When two successive readings within 1mmHg of each other are
recorded, a sound may be emitted (if sounds are enabled in the User Menu Options). If
successive readings of within 1mmHg of each other are not obtained, Keeler recommend
taking up to four readings.
9. When two consecutive readings are within 1mmHg an audible sound will be heard
indicating that sufficient readings may have been taken.
10. If a reading is recorded as a non-event or bad event, a long high pitch tone will be
heard.
11. The first reading will be the measured value; successive readings will display the running
average IOP. Outlying or spurious readings will automatically be excluded from the
calculation.
12. At any time pressing the Review button will
allow you to view the individual readings.
13. If the unit does not fire, repeat step 3-7.
To measure the other eye, press the OD / OS
button on the Hand Unit and repeat processes
3-7.
Note:
For quick alignment reference, please refer to
the short form instructions located at the rear of the
instrument and accessed by using the pull out tab on
the back left hand side.
Summary of Contents for Pulsair IntelliPuff
Page 1: ...Non Contact Tonometer INSTRUCTIONS FOR USE EN EL PL NL HU...
Page 2: ......
Page 4: ......
Page 29: ...25 Pulsair Intellipuff KEELER EL Keeler Keeler EP29 32777 Pulsair IntelliPuff Keeler Keeler...
Page 30: ...26 Pulsair Intellipuff KEELER EL Pulsair IntelliPuff 1 2 3 IOP IOP 3 3 1 1 2 3 4...
Page 33: ...29 Pulsair Intellipuff KEELER EL 4 5 6 Keeler 7 8 OD OS OD OS 2 1 3 4 5 6 7 8 14 10 12 13 11 9...
Page 39: ...35 Pulsair Intellipuff KEELER EL 10 10 1 20 IOP xx x xx 11 1 2 3 4 12 1 3 2 3 2 1...