A single reading can sometimes be misleading as the IOP will vary because of pulse, respiratory
and diurnal fluctuations. In addition, blinking, squeezing, fluid intake, physical activity, body
position and even the direction of gaze can influence IOP.
Up to 4 readings may be required to reduce the impact of these variants to obtain a constant IOP.
Pulsair IntelliPuff Tonometer software will recognise the readings and emit a sound notification
when two consecutive readings are +/-1mmHg of each other indicating that further
measurements may not be required.
7. NAMES OF CONTROLS AND COMPONENTS
1
On / Off Push Button
To turn the Pulsair IntelliPuff on or off, push the On / Off button – a green LED will indicate the
unit is on.
2
Test Eye
This is useful for user training; it will not return an IOP measurement.
3
Printer Active LED
When lit, this indicates the printer is activated; replacing the handset in the cradle will start the
print. Alternatively, a print can be made at any time using the print button on the handset.
Remove labels