Intellipuff Service Manual
Page 29 of 51
Issue D 27/02/2014
EP59-12317
Pin Number Colour
Black
White
Yellow
Red
Orange
Blue
Green
Brown
Turquoise
Purple
Pink
Grey
Fit new cable ties to retain the cableform as originally fitted.
Refit the air tube to the air filter (see changing air filter)
Whilst the handset assembly has been calibrated as part of the test procedure
it is recommended that the calibration is checked in accordance with the
instructions under Calibration including checking the air pressure relief valve
setting.