![KBR multicount 3D6-0-LCD-EP-US1 Operating Instructions, Technical Parameters Download Page 35](http://html1.mh-extra.com/html/kbr/multicount-3d6-0-lcd-ep-us1/multicount-3d6-0-lcd-ep-us1_operating-instructions-technical-parameters_659463035.webp)
User Manual
multicount
Version 2.00
Page 29 of 34
ED
EBD
A
014
5
/ 47
09
-1
GB
8.4
Measuring principle
8.5
Device memory
8.6
Power supply
8.7
Hardware inputs
Reading
128 points per period at 50Hz
Update time
~ 500ms
Frequency range
Automatic frequency adjustment between 40-70 Hz
Main data and program memory
1MB RAM battery-buffered/ 256k EPROM
Memory type
Ring buffer
Long-term memory (1 year)
(
multicount 3D6-1-LCD-ES-...
only)
Annual energy memory daily values for active and reactive energy, for high and
low tariff, for consumption and recovery for one year
Long-term memory for max. 160 days, min. 64 hours
depending on memory configuration
(
multicount 3D6-1-LCD-ES-...
only)
Load profile memory: A maximum of 4*3840 entries; 60 / 30 / 15 / 1 – minute
average values of P
Σ
and Q
Σ
for high and low tariff, for consumption and
recovery; configurable via operating software!
Long-term memory (1 year)
(
multicount 3D6-1-LCD-ES-...
only)
Measuring period maximum Pkum / Qkum per month for 1 year
Event memory
(
multicount 3D6-1-LCD-ES-...
only)
A maximum of 4096 entries to record tariff switching commands, mains failures,
error messages, and many more.
Parameter memory
non volatile
Password memory
(
multicount 3D6-1-LCD-ES-...
only)
Code 4-digit
Power supply
85 to 265V AC/DC; 50/60Hz
Power consumption
15 VA
Measuring input for
voltage
U
L1-L2
; U
L2-L3
; U
L3-L1
3 * 5V ...100V...120V AC (measuring range 1)
3 * 20V...500V...600V AC (measuring range 2)
Direct impedance
1.2 MOHM (Ph-Ph)
Measuring range
programmable
Measuring inputs for
current
I
L1
; I
L2
; I
L3
3 * 0,01A...1A...1,2A AC (measuring range 1)
3 * 0,05A...5A...6A AC (measuring range 2)
Power consumption
≤
0,3VA per input at 6A
Measuring range
programmable
Digital inpputs
(
multicount 3D6-
1-LCD-ES-...
only)
Tariff input
digital input for floating contact
Switching HT/LT, signal e.g. from energy supplier
Contact open => Tariff HT
Contact closed => Tariff LT
Synchronous input
digital input for floating contact
Synchronization of measuring period; pulse length
≥
250ms
Power supply
27V / 15mA DC
Summary of Contents for multicount 3D6-0-LCD-EP-US1