Guidelines and Manufacturer Declaration – Electromagnetic Interference Immunity
The KaWe MASTERLIGHT® HL / LED is designed to be used in the types of environments listed below.
The customer or user of the KaWe MASTERLIGHT® HL / LED is responsible for ensuring that this device is used in such an environment.
Interference immunitytest IEC 60601-test level
Compliance level
Electromagnetic environment
– guidelines
Electrostatic
discharge (ESD)
according to
IEC 61000-4-2
± 6kV contact discharge
± 8 kV air discharge
± 6kV contact discharge
± 8 kV air discharge
The flooring should be wood
or concrete or be covered with
ceramic tile. If the floor material is a
non-conductive, synthetic material,
the relative humidity of the air must
be at least 30%.
Fast transient electrical
disturbances/ bursts according
to IEC 61000-4-4
± 2kV for power supply lines
± 1kV for input and
output lines
=> not applicable
± 2kV for power supply lines
± 1kV for input and
output lines
=> not applicable
The quality of the mains supply
voltage should be the same as that
of a typical commercial or hospital
environment.
Surges according to
IEC 61000-4-5
± 1 kV differential mode
voltage
± 2 kV common mode voltage
± 1 kV differential mode
voltage
± 2 kV common mode voltage
The quality of the mains supply
voltage should be the same as that
of a typical commercial or hospital
environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
according to IEC 61000-4-11
< 5 % UT
(>95 % dip in the UT )
for a ½ period
40 % UT
(60 % dip in the UT )
for 5 periods
70 % UT
(30 % dip in the UT )
for 25 periods
< 5 % UT
(>95 % dip in the UT )
for 5 seconds
< 5 % UT
(>95 % dip in the UT )
for a ½ period
40 % UT
(60 % dip in the UT )
for 5 periods
70 % UT
(30 % dip in the UT )
for 25 periods
< 5 % UT
(>95 % dip in the UT )
for 5 seconds
The quality of the mains supply
voltage should be the same as that
of a typical commercial or hospital
environment.
If the user of the KaWe MASTER-
LIGHT® HL / LED requires continued
operation during power interrup-
tions, it is recommended that the
KaWe MASTERLIGHT® HL / LED be
powered from an uninterruptable
power supply or a battery.
Magnetic field with a power
frequency of (50/60 Hz) ac-
cording to IEC 61000-4-8
3 A/m
30 A/m
Magnetic fields with their line
frequencies should be the same as
those typical of commercial and
hospital environments.
Note:
Uτ is the mains voltage prior to application of the test level.