Instructions for use SONICflex paro
6 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 6.2 Manual Reprocessing
17 / 24
6.2 Manual Reprocessing
For effective reprocessing, automated internal and external cleaning as well as
automated internal and external disinfection with a cleaning and disinfection
unit in accordance with EN ISO 15883-1 is required.
This product is not designed for manual internal and external cleaning and
manual internal and external disinfection.
6.3 Automated reprocessing
CAUTION
Never reprocess the medical device with chloride-containing products.
Malfunction and material damage.
4
Reprocess it in a washer disinfector only.
CAUTION
Never reprocess this medical device in an ultrasonic cleaner.
Malfunction and material damage.
4
Reprocess it in a washer disinfector only.
6.3.1 Pre-cleaning
Accessories required:
▪ Tap water (drinking water)
▪ Brush, e.g. medium-hard toothbrush
4
Brush off under flowing tap water at least 10 seconds.
4
If necessary, clear the water passage at the tip with a nozzle needle.
6.3.2 Automated internal and external cleaning and
internal and external disinfection
NOTE
For reprocessing, place SONICflex tips on the cleaning adapter.
Material summary
Material number
Cleaning adaptor for SONICflex tips,
long thread
1.009.7875
Cleaning adapter for SONICflex tips,
short thread
1.009.7876
①
Cleaning adapter
②
SONICflex Apex
4
Use the wrench to the screw the SONICflex tip
②
with the aid of a torque
wrench onto the cleaning adapter
①
in a clockwise direction.
4
Remove the torque wrench.