Instructions for use SONICflex paro

2 Safety | 2.1 Infection hazard

7 / 24

2 Safety

NOTE

All serious events occurring in relation to the product must be reported to the
manufacturer and the competent authority of the member state, in which the
user and/or patient resides.

The instructions for use are an integral part of the product and must be read
carefully prior to use and must be accessible at all times. 
The device may only be used in accordance with the intended use, any other
type of use is not permitted.

2.1 Infection hazard

Patients, users or third parties could be infected by contaminated medical
devices.

4

Take suitable personal protective measures.

4

Follow the instructions for use of the components.

4

Before initial startup and after each use, reprocess the product and ac-
cessories appropriately.

4

Carry out the reprocessing as described in the instructions for use. The pro-
cedure has been validated by the manufacturer.

4

If you deviate from this validated procedure, make sure that the repro-
cessing procedure is effective.

4

Reprocess the product and accessories appropriately before disposal.

4

In the case of injury to soft tissue, do not continue treatment in the oral
cavity with instruments driven by compressed air.

4

Place the torque wrench on the tip when the handpiece is in its holder.

2.2 Technical condition

A damaged product or damaged or NOT KaVo original components could injure
patients, users or third parties.

4

Only operate devices or components if they show no signs of damage on the
outside.

4

Check to make sure that the device is working properly and is in satisfactory
condition before each use.

Using the product at improper or excessive power settings can lead to tip
breakage and ensuing injury.

4

Do not select improper or excessive power settings.

4

Check the tip for secure attachment before treatment.

Fracture may occur as a result of long-term use or damage (dropped to the
floor or mechanical change of the original shape).

4

 Before each use, check the operational safety of the tips by gently pressing
on them with your thumb or forefinger.

4

In addition, expose the tips to approx. 10 N (1 kg) mechanical load without
function.

Worn tips can fracture or be contaminated and thus cause injury or infection.

4

Due to the natural wear and tear of consumables, there is no warranty on
SONICflex tips. Exchange SONICflex tips every 9-12 months.

4

Observe the tip wear and tear measuring card.

4

Have parts with sites of breakage or surface changes checked by the service
personnel.

4

Safety checks may only be performed by trained service personnel.

Summary of Contents for SONICflex paro

Page 1: ...x paro tip no 60 0 571 7402 SONICflex paro tip no 60 A 1 006 1934 SONICflex paro tip no 61 0 571 7412 SONICflex paro tip no 61 A 1 006 1935 SONICflex paro tip no 62 0 571 7422 SONICflex paro tip no 62...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 88400 Biberach Germany www kavo com...

Page 3: ...nd storage conditions 10 4 Startup and shut down 11 4 1 Inserting the SONICflex tips 11 4 2 Removing the SONICflex tip 12 5 Operation 13 5 1 Setting the power on the SONICflex 13 5 2 Notes for applica...

Page 4: ...ical Service If you have any technical questions or complaints please contact the KaVo Technical Service 49 0 7351 56 1000 service instrumente kavo com Target group The instructions for use are intend...

Page 5: ...C Code UDI symbol CE mark for medical devices EAC conformity mark Eurasian Conformity Medical device labelling of medical devices Transportation and storage conditions temperature range Transportation...

Page 6: ...nd material damage The warning notes are designated as shown below HAZARD In cases which if not prevented directly lead to death or severe injury WARNING In cases which if not prevented can lead to de...

Page 7: ...vity with instruments driven by compressed air 4 Place the torque wrench on the tip when the handpiece is in its holder 2 2 Technical condition A damaged product or damaged or NOT KaVo original compon...

Page 8: ...with the product by the manufacturer 4 Only use accessories that have been approved for combination with the product by the manufacturer 4 Do not make any modifications to the device unless these hav...

Page 9: ...e applicable accident prevention regulations And to use these instructions for use by the qualified user According to these regulations the user is required exclusive use of equipment that is operatin...

Page 10: ...KaVo SONICflex 2003 KaVo SONICflex 2000 Tips with short thread Tips with A can be used for all versions of KaVo SONICflex quick 2008 3 5 Transportation and storage conditions CAUTION Startup after re...

Page 11: ...product and ac cessories Also refer to 6 Reprocessing steps in accordance with ISO 17664 1 ISO 17664 2 Page 16 Currently applicable packaging law Dispose of and recycle the packaging appropriately in...

Page 12: ...turning it clockwise 4 Use the thin region of the handle for insertion 4 Use the thick region of the handle for tightening or detachment NOTE The tip is properly tightened when the torque wrench skip...

Page 13: ...el on the control ring of the SONICflex 5 2 Notes for application This SONICflex paro tips can be used for initial treatment and regular follow up treatments Regular follow up treatments are very impo...

Page 14: ...n any quadrant The angled SONICflex paro tips can be used in pockets that are difficult to access If you hold the tip by the thread to look at it you can see a slight angling to the right or left The...

Page 15: ...l Mandible right vestibular Maxilla right lingual Maxilla left vestibular CAUTION Damage to the root surface Damage to the root surface from placing the front of the tip against the root surface 4 Onl...

Page 16: ...wear protect ive gloves 4 Reprocess the medical device right after treatment 4 Remove all residual cement composite or blood immediately 4 Wipe disinfect the medical device before transport 4 Remove t...

Page 17: ...TION Never reprocess this medical device in an ultrasonic cleaner Malfunction and material damage 4 Reprocess it in a washer disinfector only 6 3 1 Pre cleaning Accessories required Tap water drinking...

Page 18: ...r use of the washer disinfector 4 In order not to affect the KaVo medical device make sure that the product is dry on the inside and outside after completion of the cycle 6 4 Care products and systems...

Page 19: ...least 30 minutes at 121 o C 1 o C 4 o C 250 o F 1 6 o F 7 4 o F 4 Remove the medical device from the steam steriliser immediately after the sterilisation cycle is completed 4 Use according to the man...

Page 20: ...lers Consumables Material summary Material number Torque wrench 1 000 4887 Nozzle needle 0 410 0911 Tip wear and tear measuring card 1 001 6958 Cleaning adapter Material summary Material number Cleani...

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ...0 589 3072 s 20231129 03 en...

Reviews: