Instructions for use SONICflex paro
6 Reprocessing steps in accordance with ISO 17664-1 / ISO 17664-2 | 6.7 Storage
19 / 24
▪ Steriliser using the gravity method:
– at least 10 minutes at 134
o
C -1
o
C / +4
o
C (273
o
F -1.6
o
F / +7.4
o
F)
– at least 30 minutes at 121
o
C -1
o
C / +4
o
C (250
o
F -1.6
o
F / +7.4
o
F)
4
Remove the medical device from the steam steriliser immediately after the
sterilisation cycle is completed.
4
Use according to the manufacturer's Instructions for Use.
6.7 Storage
Reprocessed products must be stored appropriately protected from dust in a
dry, dark and cool space with a low germ level.
NOTE
Comply with the expiry date of the sterilised items.