Factory presets for image resolution High Definition (HD)
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
90
5
4.5
Small adult (S)
90
6.3
5.6
Medium-sized adult (M)
90
8
7.1
Large adult (L)
90
10
9
Extra large adult (XL)
90
11
11
Factory presets for image resolutions Hi Res and Endo
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
90
6.3
5.6
Small adult (S)
90
8
7.1
Medium-sized adult (M)
90
10
9
Large adult (L)
90
11
10
Extra large adult (XL)
90
11
11
If you need to adjust the preset exposure values for this exposure:
1. Select the kV / mA field.
2. Use the minus or plus buttons to set the exposure values you wish to
use. To improve the image contrast, reduce the kV value. To reduce the
radiation dose, reduce the mA value.
3. Select the green check mark button.
NOTE
You can adjust the preset exposure values permanently as described in
section "Programs (2100)" on page 62.
8.6 Selecting patient movement correction
Steps
1. Use this button to select the CALM patient movement correction
function.
This is an algorithm that detects patient movement during exposure
and then compensates for the effects of the movement during image
reconstruction.
8 3D patient exposure
Instructions for use
KaVo ProXam 41