NOTE
If you have any reason to believe that the C-arm might hit the patient
during exposure (e.g. patients with wide shoulders), take a test exposure
without radiation first. To switch radiation off, select Settings > User > 1300
Operational settings > 1310 Use Mode > 1311 Set Demo Mode.
Alternatively, use the 90° button at the bottom of the screen to test the
C-arm rotation before you take a 3D exposure.
NOTE
Do not touch the arm structures when the X-ray unit is moving.
NOTE
Patients are not allowed to hang on the patient handles.
NOTE
FOR X-RAY UNITS WITH ROTATING TUBE HEAD:
Do not touch the metal parts in the middle of the tube head. They might be
hot.
NOTE
Portable mobile devices and other high frequency electromagnetic energy
emitting devices used close to the X-ray system may affect the system’s
performance. Diagnostic information of the X-ray image may be lost and
unnecessary X-ray dose to the patient may result.
4.1 Reporting serious incidents
Serious incidents that have occurred in relation to the device must be
reported to the manufacturer and the local competent authority.
4 Safety precautions
8 KaVo ProXam
Instructions for use