The exposure values have been preset at the factory for each patient size,
image resolution and ULD (Ultra Low Dose) setting. The preset exposure
values are average values and they are only meant to guide the user.
The preset exposure values are shown in the following tables.
Factory presets for image resolution Low dose in programs Nose and Sinus
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
90
2.8
1.4
Small adult (S)
90
3.6
1.8
Medium-sized adult (M)
90
5.0
2.2
Large adult (L)
90
5.6
2.8
Extra large adult (XL)
90
7.1
3.6
Factory presets for image resolution Low dose in other programs
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
90
4
2
Small adult (S)
90
5
2.5
Medium-sized adult (M)
90
6.3
3.2
Large adult (L)
90
8
4
Extra large adult (XL)
90
10
5
Factory presets for image resolution Normal in programs Nose and Sinus
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
90
3.2
2.2
Small adult (S)
90
4
2.8
Medium-sized adult (M)
90
5
3.6
Large adult (L)
90
6.3
4.5
Extra large adult (XL)
90
8
5.6
Factory presets for image resolution Normal in other programs
Patient size
kV value
mA value
mA value with
ULD
Child (XS)
90
5
3.6
Small adult (S)
90
6.3
4.5
Medium-sized adult (M)
90
8
5.6
Large adult (L)
90
10
7.1
Extra large adult (XL)
90
11
9
8 3D patient exposure
40 KaVo ProXam
Instructions for use