4 Commissioning | 4.4 Connecting the surgical motor
24 / 76
4.4 Connecting the surgical motor
Note
The delivered parts are not sterile (except for the coolant hose). Before the first
treatment of a patient, the surgical motor, motor cable, and the handpiece tray need
to be reprocessed.
See also:
2
7 Reprocessing steps in accordance with DIN EN ISO 17664, Page 58
1
2
3
4
① Motor coupling
② Surgical motor
③ Handpiece tray
④ Plug of motor cable
▶ Plug the surgical motor ② into the motor coupling ① and secure it with a union
nut. Please note the separate instructions for use of the motor.
▶ Place the surgical motor on the handpiece tray ③.
▶ Insert the plug of the motor cable ④ into the connector on the device, align the
marker points, and insert the plug until it snaps into place.
4.5 Connecting the coolant container and hose set
CAUTION
Running, open hose pump.
Risk of injury.
▶ Turn off the device before opening the hose pump.