Instructions for use MASTERsurg LUX Wireless REF 1.009.1200
5 Operation | 5.7 Rinsing function
The rinsing function serves to feed liquid and to start-up the illumination on the hand-
piece. The motor is not activated during this process. The rinsing function can be
called up manually at any time.
▶ Press the pump key on the wireless foot control for an extended time to activate
the rinsing function.
▶ Press the speed key on the wireless foot control to start the rinsing function and
control the coolant flow.
▶ To terminate the rinsing function, press the back key or the pump key.
5.7.2 Program activity rinsing function
The rinsing function serves to feed liquid and to start-up the illumination on the hand-
piece. The motor is not activated during this process. The rinsing function can be
defined as an activity in the course of the program.
▶ Defining and executing program steps.
See also:
2
Example 2: Program steps that do not include the activity, "Cut thread", and in-
clude the "Rinsing function" activity, Page 42
49 / 76