Instructions for use MASTERmatic LUX M10 L - 1.009.3570
3 Product description | 3.2 Technical Specifications
3.2 Technical Specifications
Drive speed
max. 40,000 rpm
-1
Transmission ratio
1:1
Maximum speed:
max. 40,000 rpm
-1
Identification
1 blue ring
Spray water pressure
0.8 to 2.0 bar (12 to 29 psi)
Spray air pressure
1.0 to 2.0 bar (15 to 29 psi)
Spray air quantity
min. 1.5 Nl/min (at 2 bar)
Cooling air flow
5.5 to 9.5 Nl/min
Handpiece cutters or grinders can be used.
Handpiece cutters or grinders can be used.
The handpiece can be mounted on all INTRAmatic (LUX) motors, and motors with a
connection in accordance with ISO 3964 / DIN 13940.
3.3 Transportation and storage conditions
CAUTION
It is hazardous to start up the medical device after it has been stored strongly refriger‐
ated.
This can cause the medical device to malfunction.
▶ Prior to start-up, very cold products must be heated to a temperature of 20
o
C to
25
o
C (68
o
F to 77
o
F).
Temperature: -20°C to +70°C (-4°F to +158°F)
Relative humidity: 5% RH to 95% RH absence of condensation
Air pressure: 700 hPa to 1060 hPa (10 psi to 15 psi)
Protect from moisture