KaVo MASTERmatic LUX M10 L Instructions For Use Manual Download Page 23

Instructions for use MASTERmatic LUX M10 L - 1.009.3570

8 Tools and consumables | 7.8 Storage

8 Tools and consumables

Available from dental suppliers.

Material summary

Mat. No.

Instrument stand 2151

0.411.9501

Cleanpac 10 units

0.411.9691

Cellulose pad 100 units

0.411.9862

Nozzle pin

0.410.0921

Spray hose, sterilisable

0.065.5188

Drill stop

0.524.0892

Hook

0.410.1963

Spray head INTRA (KaVo Spray)

0.411.9911

Material summary

Mat. no.

KaVo CLEANspray 2110 P

1.007.0579

KaVo DRYspray 2117 P

1.007.0580

KaVo Spray 2112 A

0.411.9640

ROTAspray 2 2142 A

0.411.7520

QUATTROcare plus Spray 2140 P

1.005.4525

 

23 / 26

Summary of Contents for MASTERmatic LUX M10 L

Page 1: ...Instructions for use MASTERmatic LUX M10 L 1 009 3570...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 7351 56 0 Fax 49 7351 56 1488 Manufacturer KaVo Dental GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...6 2 2 Cleaning the spray nozzle 14 7 Reprocessing steps in accordance with ISO 17664 16 7 1 Preparations at the site of use 16 7 2 Cleaning 16 7 2 1 Manual cleaning external 16 7 2 2 Automated externa...

Page 4: ...9 Warranty terms and conditions 24 Instructions for use MASTERmatic LUX M10 L 1 009 3570 Table of contents 4 26...

Page 5: ...Symbols Refer to the chapter on Safety Warning symbol Important information for users and service technicians Can be steam sterilised at 134 o C 1 o C 4 o C 273 o F 1 6 o F 7 4 o F Thermodisinfectable...

Page 6: ...and injury CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moder ate injuries WARNING WARNING indicates a hazardous situation that can lead to serious or...

Page 7: ...y and performance The medical device may only be combined with a treatment centre released by KaVo CAUTION Premature wear and malfunctioning from improper storage during long periods of nonuse Reduced...

Page 8: ...rations removal of fillings processing of tooth and restoration surfaces A medical device according to relevant national statutory regulations Proper use According to these regulations this medical de...

Page 9: ...can be used Handpiece cutters or grinders can be used The handpiece can be mounted on all INTRAmatic LUX motors and motors with a connection in accordance with ISO 3964 DIN 13940 3 3 Transportation a...

Page 10: ...te manner Prior to disposal the product and accessories must be appropriately prepared or sterilised if this is necessary 4 1 Check the amount of water CAUTION Overheating of the tooth due to insuffic...

Page 11: ...ece while the device is rotating Lightly spray O rings on motor coupling with KaVo Spray Attach medical device to the motor coupling and turn it until you hear the latch snap into place Pull on the me...

Page 12: ...rely held When checking inserting and removing use gloves or a fingerstall to prevent an injury or infection Rotate the clamping ring all the way in the direction of the arrow and insert the handpiece...

Page 13: ...10 L 1 009 3570 5 Operation 5 5 Conversion for contra angle handpiece drill bit Press the contra angle drill bit onto the stop close the clamping ring and check for firm seating To remove the drill bi...

Page 14: ...the medical device An O ring is missing on the motor coupling Replace O ring 6 2 Troubleshooting 6 2 1 Replacing the O rings CAUTION Hazard from improper care of the O rings Malfunctions or complete...

Page 15: ...Instructions for use MASTERmatic LUX M10 L 1 009 3570 6 Troubleshooting 6 2 Troubleshooting Use the nozzle needle Mat no 0 410 0921 to free the water passage in the spray nozzles 15 26...

Page 16: ...asonic unit Defects in the device Only clean manually or in a thermodisinfector 7 2 1 Manual cleaning external Accessories required Tap water 30 o C 5 o C 86 o F 10 o F Brush e g medium hard toothbrus...

Page 17: ...15883 1 which are operated using alkaline cleaning agents For programme settings as well as cleansers and disinfectants to be used please refer to the Instructions for Use of the washer disinfector In...

Page 18: ...ve been released by KaVo with respect to the compatibility of materials e g WL cid made by ALPRO Blow off with compressed air until no water drops are visible Immediately after internal disinfection l...

Page 19: ...our products 7 5 1 Care with KaVo Spray KaVo recommends servicing the product as part of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the c...

Page 20: ...ore each sterilisation Remove the cutter or grinder Servicing the product See also 2 Instructions for use KaVo QUATTROcare 2104 2104A Servicing of the clamping chuck KaVo recommends cleaning and servi...

Page 21: ...gainst the tip of the service coupling Press the button marked with the chuck service symbol Note Close the chuck service mode Option 1 Place the dental handpieces in the QUATTROcare PLUS 2124 A close...

Page 22: ...Before each sterilisation cycle service the medical device with KaVo care prod ucts CAUTION Contact corrosion due to moisture Damage to product Immediately remove the product from the steam sterilise...

Page 23: ...9501 Cleanpac 10 units 0 411 9691 Cellulose pad 100 units 0 411 9862 Nozzle pin 0 410 0921 Spray hose sterilisable 0 065 5188 Drill stop 0 524 0892 Hook 0 410 1963 Spray head INTRA KaVo Spray 0 411 99...

Page 24: ...consequences that have arisen or may arise from to natural wear improper handling cleaning or maintenance non compliance with operating maintenance or connection instructions calcination or cor rosio...