2 Safety | 2.2 Safety instructions
CAUTION
Risk due to incorrectly stored instrument.
Injury and infection caused by chucked cutters or grinders.
Damage to clamping system from dropping the instrument.
▶ After treatment, place the instrument properly in the cradle, without the cutter or
grinder.
CAUTION
Hazard from use as a light probe.
Do not use the device as a light probe since the rotating cutters or grinders can cause
injury.
▶ Use an appropriate light probe for additional illumination of the oral cavity or site
of preparation.
CAUTION
Risks due to lack of control equipment.
Hazards can arise if control equipment is not available for changing the speed and
the direction of rotation.
▶ The dental treatment unit connected must have control equipment for changing
the speed and direction of rotation.
▶ A note is to be included in the documents accompanying the dental treatment
unit, referring to responsibilities arising from safety, reliability and performance.
▶ The medical device may only be combined with a treatment centre released by
KaVo.
CAUTION
Premature wear and malfunctioning from improper storage during long periods of
nonuse.
Reduced product life.
▶ The medical device should be cleaned, serviced and stored in a dry location, ac‐
cording to instructions, before long periods of nonuse.
Note
For safety reasons, we recommend that the tool holder system be checked annually
after the warranty period expires.
The following individuals are authorized to repair and service KaVo products:
▪ Technicians at KaVo branches throughout the world
▪ Technicians specially trained by KaVo
To ensure proper function, the medical device must be set up according to the reproc‐
essing methods described in the KaVo Instructions for Use, and the care products and
care systems described therein must be used. KaVo recommends specifying a service
interval at the dental office for a licensed shop to clean, service and check the func‐
tioning of the medical device. This service interval depends on the frequency of use
and should be adjusted accordingly.
Service may only be carried out by KaVo-trained repair shops using original KaVo re‐
placement parts.