Instructions for use MASTERmatic LUX M10 L - 1.009.3570
7 Reprocessing steps in accordance with ISO 17664 | 7.5 Care products and systems - Servicing
7.5 Care products and systems - Servicing
WARNING
Sharp cutters or grinders in the medical device.
Risk of injury from sharp and/or pointed cutters or grinders.
▶ Remove cutter or grinder.
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶ Perform proper care regularly!
Note
KaVo only guarantees that its products will function properly when the care prod‐
ucts are used that are listed as accessories since they were tested for proper use
on our products.
7.5.1 Care with KaVo Spray
KaVo recommends servicing the product as part of the reprocessing after each use,
i.e. after each cleaning, disinfection, and before each sterilisation.
▶ Remove the cutter or grinder.
▶ Cover the product with the Cleanpac bag.
▶ Plug the product onto the cannula, and press the spray button for one second.
Servicing of the clamping chuck
KaVo recommends cleaning and servicing the chuck system once a week.
▶ Remove the cutter or grinder, place the spray nipple tip in the opening and spray.
▶ Carry out the servicing according to the instructions in the section, "Care with Ka‐
Vo Spray".
7.5.2 Care with the KaVo SPRAYrotor
▶ Place the product on the appropriate coupling on the KaVo SPRAYrotor and cov‐
er it with the Cleanpac bag.
▶ Servicing the product.
See also:
2
Instructions for use KaVo SPRAYrotor
19 / 26