Dear dentist:
KaVo hopes that you will benefit greatly from working with your new quality product, the
INTRA LUX motor KL 701.
Please observe the following notes to assure continued trouble-free, economical and safe product
operation:
Technical Data
INTRA LUX motor KL 701
with motor electronics,
mat.-no. 0.696.0161
Motor voltage
min. 1 V DC
max. 22 V DC
Motor speed
min. 2,000 rpm
max. 40.000 rpm
Motor torque
max. 2,7 Ncm
Contact the manufacturer for connected loads at the
device.
Mode: intermittent service
2 min., 5 min. pause
High-pressure lamp
operating voltage
max. 3.2 V DC
Power of the high-pressure lamp,
max. 2.5 W
Cooling air
2.5 - 4.5 bar (36-64p.s.i)
Quantity of cooling air
20 ± 2 NL/min.
Exit of air on motor coupling
7.5 ± 1,5 NL/min.
Spray air
1.0 - 2.5 bar (14-36 p.s.i)
Spray water
0.8 - 2.0 bar (11-28 p.s.i)
Handpiece connection according to ISO 3964
Important Remarks
The motor is a dental low
- voltage electrical motor of type 2 according to ISO 11498
- is not admitted for use in areas with risk of explosions.
- only to be used by a qualified dental practitioner, in connection with the provision
of dental treatment to dental patients, any other use may prove dangerous and is
strictly forbidden,
- a medical product according to the applicable national provisions of law.
In conformity with these regulations, the motor may only be used by a competent
person for the described application, and under consideration of
- the valid regulations on work safety,
- the valid accident prevention measures,
- these operating instructions.
In conformity with these regulations it is the duty of the user
- to use only faultless tools,
- to ensure that it is used for the correct purpose,
- to protect patients and third parties from all dangers.
Risks from electromagnetic fields (cardiac pacemakers)
When treating patients with cardiac pacemakers or comparable implanted
systems, it must be expected that electromagnetic fields will influence the
functioning of the implanted systems. Before the beginning of the
treatment, the user must question the patient accordingly.