A 5 Installation and Connection
A 5.1 Installation
The following are authorized to
repair and service the KaVo
ELECTROtorque plus 4892:
• the engineers of KaVo agents through-out
the world
• the engineers of authorised KaVo dealers
who have been specially trained by KaVo
• independent engineers specially trained
by KaVo.
• Before opening housing parts, pull out
the mains plug or completely isolate
from the mains connection and thus cut
off all power to the unit.
Location at the dental unit:
Place the ELECTROtorque plus 4892 in the
intended location (see also A 4.1).
Remove previously used turbine or
pneumatic motor from tube and connect
turbine tube or pneumatic motor tube
connection
4
to four-hole connector
3
at
the back of the ELECTROtorque plus
4892.
Location of transformer:
Connect the ELECTROtorque plus 4892 to
the intended transformer
5
via screw
connection
2
.
Factory settings for the transformer
5
is 230 V,
Location of INTRAmatic LUX 700 KL /
701 KL:
Mount motor INTRAmatic LUX 700 KL /
701 KL on tube
1
and place in holder of
dental unit.
A 5.2 Voltage adaptation to
Transformer 1.001.8749
By removing the fuse holder
7
out of the letting modul
8
,the bridge
6
can be removed.
Replace the bridge that way, that the
relected voltage (115 or 230 V) can be
read.
Finally please replase the fuse holder
7
into the setting modul
8
.
ELECTROtorque plus 4892.