9
ELECTROtorque plus 4892.
A 3 Technical data
A 3.1 ELECTROtorque plus 4892
Dimensions and weights
width:
5 inches
depth:
6 inches
height:
2 inches
weight:
780 grams
Ambient conditions:
Permitted indors
Permitted ambient temperature range of
41°F - 104°F (5°C - 40°C)
Permitted relative humidity up to max. 80%
mode of operation
intermittent operation
1 min.
pause
9 min.
protection class:
II
pieza del uso:
Type B
protection art:
IP
40
Air Requirements:
dry, free from contamination and oil
if necessary
• Use a compressor with a dry air system.
• Install an air filter if necessary (50 µm).
• Blow out the lines before installation.
• system pressure :
1,8 - 4,5 bar
• air flow at e-type connection:
6 - 10 NL / min
spray air pressure:
1,0 - 2,5 bar
spray water pressure:
0,8 - 2,0 bar
Ideal settings at dental unit
- system pressure:
3 bar
- spray air pressure:
1 bar
- spray water pressure:
0,8 bar
We reserve the right to make technical
modifications.
150
55
125
x10
00 r
pm
ELE
CTR
Oto
rque
plu
s
S1
S2
rever
se
x10
00 r
pm
ELE
CTR
Oto
rque
plus
S1
S2
reve
rse