A 6.3 Setting the lamp voltage
The lamp voltage can be changed in the
range from 3.0 V to 3.8 V (measured at the
housing plug) with the motor switched off.
The voltage at the lamp itself is 0,5 V
below the set value.
The factory setting on delivery is 3.2 V
(about 2.7 V at the lamp).
Call up the “Lamp voltage setting“
subroutine by pressing (more than 1 seconds)
the keys “REVERSE“
6
and
1
↑
together. The display
3
shows the value
last saved (in volt).
This value last saved can be changed
between 3.0 V and 3.8 V by pressing the
keys
1
↑
and
2
↓
.
The setting is completed by pressing key
“REVERSE“
6
for a long time, and the
setting is saved.
A 6.4 Set lamp delay
The lamp delay can be adjusted from
0 seconds to 10 seconds in steps of 1 seconds.
The facttory setting is 3 seconds.
By pressing “REVERSE“
6
and S2
4
simultaneity (more than 1 seconds) call
subroutime “lamp delay“. The display
shows the current value of lamp delayin
seconds.
Now this value can be changed by pressing
1
↑
und
2
↓
.
Save the selected value by pressing
“REVERSE“
6
for 2-3 seconds.