Instructions for use HealOzone 2130 C
9 Safety checks
|
9.2 Method
9.2 Method
9.2.1 Earth wire impedance measurement
The following thresholds must not be exceeded: < 0.3 Ω with mains cable and < 0.2
Ω without.
▶ Using a test probe, sample the rear panel of the casing.
9.2.2 Replacement device leakage current measurement
Must not fall below the < 1 mA threshold.
L
N
MD
~
56/57