Instructions for use HealOzone 2130 C
5 Operation
|
5.2 The treatment procedure
If the measured moisture level lies above the range of the assigned moisture level,
this status is shown in the display.
The start of drying is reported acoustically by a double buzzer tone.
▪ Treatment time
▪ Drying process running
If the flow sensor becomes inactive during the drying process, the unit displays
message "Waiting for treatment". You can choose the Stop button to cancel and it
will switch directly to rinse mode.
5.2.8 Treatment
An airtight seal has been created. Ozone is produced from this point on. The unit
counts down the seconds for the remainder of the treatment time. Once the treat‐
ment time has come to an end, the unit switches to rinse mode. It is no longer
possible to change the treatment time.
You can press the stop button or the footswitch to stop the process. If the process
is stopped, the unit switches directly to rinse mode.
Screen information:
▪ Remaining treatment time
▪ Ozone application
5.2.9 Ozone generation
If it takes a while to generate ozone, "Generate ozone" is shown on the display and
reported acoustically with a double buzzer tone.
Screen information:
▪ Treatment time
▪ Ozone is being generated
Once the ozone starts being generated, the device switches immediately to treat‐
ment status.
If ozone cannot be generated within 60 seconds, the warning "Ozone generation
not successful" is output.
The message can be acknowledged with any button, and the treatment process can
be restarted.
5.2.10 Vacuum lost
If the airtight seal over the tooth is broken during treatment, ozone production is
stopped and air is blown through the ducts to clear them. The unit waits until the
vacuum has been recreated. The "Seal cup" line starts flashing. The treatment time
cannot be changed. If the airtight seal cannot be recreated within 60 seconds, treat‐
ment is cancelled. A warning message appears on screen. To stop the process,
press the stop button or footswitch. If you do this, the unit switches directly to rinse
mode.