Instructions for use HealOzone 2130 C
9 Safety checks
|
9.1 Basics
9 Safety checks
9.1 Basics
Note
You are required to comply with all safety information provided in the User instruc‐
tions under section 2 Safety.
According to VDE 0751-1:
▪ Safety checks every two years
▪ Device type IIa
▪ Protection class I
▪ Portable device
▪ Type B general
▪ Measurement in accordance with EGA
KaVo supplies a medical device record in which you make an inventory listing and
record the main master data of the medical device. www.kavo.com/De/produkte/
praxis_einrichtungen/
The following readings and measurements must be documented:
▪ Visual inspection of medical device and accessories
▪ Nominal data checks on externally accessible fuses.
▪ Earthed conductor checks in accordance with VDE 0751
▪ Leakage current measurements in accordance with VDE 0751
▪ Functional test of the medical device taking into account accompanying docu‐
ments
Note
The main unit switch must be switched on during measurement.
55/57