Instructions for use HealOzone 2130 C
8 Troubleshooting
|
8.1 Warnings
Malfunction
Cause
Remedy
Please replace
Continue with button
The initial humidity reading is >
10%. Desiccant is used up.
See also: 7.2.3 Replace desiccant,
Page 48
Desiccant is not connected.
▶ Check whether desiccant is
connected.
Counter mode
Warning: 6
Desiccant used up
Please replace
Press <SET>
The allocated humidity level is not
reached within the specified timef‐
rame.
The initial humidity reading is >
20%. Desiccant is used up.
▶ Confirm change of desiccant
with SET button, and confirm
"New desiccant" with "Yes".
The counter is reset to zero.
If the change of desiccant is con‐
firmed with "Yes", the message
reappears after the next use.
See also: 7.2.2 Sensor/counter mo‐
de, Page 47
Desiccant is not connected.
▶ Check whether desiccant is
connected.
Sensor mode
Warning: 7
Desiccant used up
Please replace
Continue with button
After the current treatment
Average humidity is ≥ the 7% thres‐
hold that was set.
Desiccant is used up
▶ Confirm the change of desiccant
with the SET button.
or
▶ Continue with any button.
Counter mode
Warning: 7
Desiccant used up
Please replace
Confirm <SET>
After the current treatment:
Counter shows that the program‐
med threshold of 600 hours opera‐
ting time or 1,000 treatments has
been exceeded.
Desiccant is used up.
▶ Confirm change of desiccant
with SET button, and confirm
"New desiccant" with "Yes".
The counter is reset to zero.
If the change of desiccant is con‐
firmed with "Yes", the message
reappears after the next use.
See also: 7.2.2 Sensor/counter mo‐
de, Page 47
52/57