7.3.3
Mechanically disinfecting the exterior and interior
KaVo recommends thermodisinfectors in accordance with ISO 15883-1,
e.g. Miele G 7781 / G 7881.
(Validation was performed with the program "VARIO-TD", the cleaner "neodisher
®
mediclean", the neutraliser "neodisher
®
Z" and rinse "neodisher
®
mielclear" and
only refers to the material compatibility with KaVo products.)
▶ For program settings as well as cleansers and disinfectants to be used, please
refer to the Instructions for Use of the thermodisinfector.
▶ Directly after automated cleaning, treat the medical device with the care prod‐
ucts and systems provided by KaVo.
7.4
Drying
Manual Drying
▶ Blow off the outside and inside with compressed air until water drops are no
longer visible.
Automatic Drying
The drying procedure is normally part of the cleaning program of the thermodisin‐
fector.
Note
Follow the instructions for use of the thermodesinfector (compressed air quality
according to ISO 7494-2).
7.5
Care products and systems – Servicing
CAUTION
Sharp cutters or grinders in the medical device.
Injury hazard from sharp or pointed cutters or grinders.
▶ Remove cutter or grinder.
CAUTION
Premature wear and malfunctions from improper servicing and care.
Reduced product life.
▶ Perform proper care regularly!
Note
KaVo only guarantees that its products will function properly when the care prod‐
ucts used are those listed as accessories, as they were tested for proper use on
our products.
Instructions for use GENTLEsilence 8000BN - 1.000.5700
7 Preparation methods according to ISO 17664 | 7.4 Drying
16/19