Distributed by:
KaVo Dental GmbH
Bismarckring 39
D-88400 Biberach
Phone +49 (0) 7351 56-0
Fax +49 (0) 7351 56-1488
Manufacturer:
Kaltenbach & Voigt GmbH
www.kavo.com
Page 1: ...Instructions for use GENTLEsilence 8000BN 1 000 5700 Always be on the safe side...
Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...
Page 3: ...the O rings on the medical device 12 6 2 Cleaning the spray nozzle 12 6 3 Change the water filter 12 7 Preparation methods according to ISO 17664 14 7 1 Preparation at the site of use 14 7 2 Cleaning...
Page 4: ...of 12 months from the date of the invoice subject to the following conditions In case of justified complaints KaVo will honour its warranty with a free replace ment or repair Other claims of any natu...
Page 5: ...rised by KaVo Warranty claims will only be accepted if the product is submitted along with proof of purchase in the form of a copy of the invoice or note of delivery The dealer purchase date type and...
Page 6: ...NING WARNING indicates a hazardous situation that can lead to serious or fatal injury DANGER DANGER indicates a maximal hazard due to a situation that can directly cause death or fatal injury 2 2 Purp...
Page 7: ...ould be cleaned serviced and stored in a dry location according to instructions before long periods of nonuse CAUTION Injury or damage due to wear Irregular running noise significant vibration overhea...
Page 8: ...air and service KaVo products Technicians at KaVo branches throughout the world Technicians specially trained by KaVo To ensure proper function the medical device must be set up according to the re pr...
Page 9: ...storage conditions CAUTION It is hazardous to start up the medical device after it has been stored strongly refrigerated This can cause the medical device to malfunction Prior to start up very cold p...
Page 10: ...m min for spray cooling 4 2 Check the pressure A drive pressure of at least 2 7 to 2 9 bar 39 to 42 psi is required for operating the medical device 2 8 bar 40 psi is ideal The air consumption is appr...
Page 11: ...oupling independently snaps into place Pull on it to make sure that the medical device is securely affixed to the cou pling WARNING Release of the medical device during treatment A medical device that...
Page 12: ...cuts Wear gloves or fingerstalls Forcefully press the push button with your thumb and simultaneously insert the cutter or grinder all the way Check that the cutter or grinder is seated by pulling on i...
Page 13: ...ury Never touch soft tissue with the head or tip since it may be hot and cause a burn After treatment is over remove the cutter or grinder from the contra angle handpiece since injury and infection ma...
Page 14: ...e spray water channels and clean the spray nozzles with the nozzle needle Mat no 0 410 0931 if necessary Check the water filter and exchange if necessary Use the nozzle needle Mat no 0 410 0921 to fre...
Page 15: ...Pull back the insert slightly Place the sleeve on the insert and screw it tight clockwise Instructions for use GENTLEsilence 8000BN 1 000 5700 6 Troubleshooting 6 3 Change the water filter 13 19...
Page 16: ...l device The medical device must be dry when transported for reconditioning Do not place it in a solution or similar 7 2 Cleaning CAUTION Malfunctions from cleaning in the ultrasonic unit Defects in t...
Page 17: ...infect in a thermodisinfector or manually 7 3 1 Manually disinfecting the exterior KaVo recommends the following products based on material compatibility Mikrozid AF made by Sch lke Mayr liquid or clo...
Page 18: ...compressed air until water drops are no longer visible Automatic Drying The drying procedure is normally part of the cleaning program of the thermodisin fector Note Follow the instructions for use of...
Page 19: ...of the reprocessing after each use i e after each cleaning disinfection and before each sterilisation Remove the cutter or grinder Servicing the product KaVo QUATTROcare plus spray can KaVo recommends...
Page 20: ...80 4 F KaVo recommends for example STERIclave B 2200 2200P by KaVo Citomat K series by Getinge Autoclave with a pre vacuum for least four minutes at 134 C 1 C 273 F 1 8 F Drying time 20 min Autoclave...
Page 21: ...y 1 003 6884 Cleanpac 10 units 0 411 9691 STERIclave bags 0 411 9912 O ring 7 4 x 0 8 1 005 8152 O ring 4 4 x 0 85 1 005 8635 NSK spray head for KAVOspray 1 005 8436 KAVOspray 2112 A 0 411 9640 NSK tu...
Page 22: ......
Page 23: ...1 005 8790 kb 20081009 03 en...