Instructions for use

DIAGNOdent pen 2190

Always be on the safe side.

Summary of Contents for DIAGNOdent 2190

Page 1: ...Instructions for use DIAGNOdent pen 2190 Always be on the safe side...

Page 2: ...Distributed by KaVo Dental GmbH Bismarckring 39 D 88400 Biberach Phone 49 0 7351 56 0 Fax 49 0 7351 56 1488 Manufacturer Kaltenbach Voigt GmbH Bismarckring 39 D 88400 Biberach www kavo com...

Page 3: ...bility 11 2 4 Disposal 11 2 5 Disposal of electronic and electrical devices 11 3 Product description 13 3 1 DIAGNOdent pen 2190 13 3 2 Purpose Proper use 14 3 2 1 General 14 3 2 2 Product specific 15...

Page 4: ...6 1 Setup methods according to DIN EN ISO 17664 33 6 1 1 Preparations for cleaning 34 6 1 2 Cleaning 34 6 1 3 Disinfection 34 6 1 4 Sterilization in a steam sterilizer in compliance with DIN EN 13060...

Page 5: ...mission RI Repair instructions RK Retrofitting kit AS Assembly set EP Enclosed parts EMC Electromagnetic compatibility PI Processing instructions 1 1 2 Symbols Refer to the chapter on Safety Warning s...

Page 6: ...rranty does not cover defects and their consequences that arose or may have arisen due to natural wear improper handling cleaning or maintenance non compli ance with operating maintenance or connectio...

Page 7: ...mmediately and no later than 7 days after delivery 2 Report the damage to KaVo 3 Leave the product and packaging in the condition in which you received it 4 Do not use a damaged product Note If the re...

Page 8: ...th CMR law Chapter 5 Art 30 1 4 3 Information on the packaging Storage and transportation Note Please keep the packaging in case you need to return the product for servicing or repair The symbols prin...

Page 9: ...d in this docu ment to prevent personal and property damage CAUTION CAUTION indicates a hazardous situation that can cause damage to property or mild to moder ate injuries WARNING WARNING indicates a...

Page 10: ...ardiac pacemaker or other system implanted before you start the treatment CAUTION Blinding hazard from laser beam of laser class 1 Eye injury Do not look into the laser beam Do not open the unit WARNI...

Page 11: ...ple and the environment If you have any questions regarding proper disposal of the KaVo product please contact the KaVo branch 2 5 Disposal of electronic and electrical devices Note According to EC di...

Page 12: ...ently installed will be picked up at the office A permanently installed unit will be picked up at the curb at your address on the agreed date The owner or user of the device will have to bear the cost...

Page 13: ...5 6 7 8 10 13 3 9 16 15 17 18 Ring switch The start button Warning sign danger laser Save button Minus button Plus button Menu button LCD display Laser notification sign Battery compartment Rating pl...

Page 14: ...ational regulations and the rules of technology applicable to the startup and use of the KaVo product for the intended pur pose are to be applied and complied with This KaVo product is intended only f...

Page 15: ...ment within the context of dental medicine It is to be used in a dentist s office or dental clinic The DIAGNOdent pen 2190 is a tool that helps dentist s detect concretions or caries in teeth that hav...

Page 16: ...applied part type B For disposal information see Intended use Follow the instructions for use icon is attached on the side of the unit Manufacturer Battery name MM Month manufactured YYYY Year manufa...

Page 17: ...e 1 mW Wavelength of the laser diode 655 nm Beam strength of the infrared diode 140mW sr Wavelength of the infrared diode 850 nm 950 nm Degree of protection IP 44 Operating temperature 10 o C to 30 o...

Page 18: ...to DIN EN ISO 17664 Page 33 4 1 Insert the battery CAUTION Injury damage due to leaky battery Health damage and product damage Use leak proof batteries only Remove the battery during long downtimes Pr...

Page 19: ...Instructions for use DIAGNOdent pen 2190 4 First use 4 1 Insert the battery Insert conventional mignon LR6 alkaline cell matching the polarity 1 2 3 Shove on the grip sleeve 19 44...

Page 20: ...tacts and en ters the tooth surface If fluorescent light arises in the case of a pathological change it is evaluated At a fissure careful scanning is required since this allows very small defects to b...

Page 21: ...onds until the DIAGNOdent pen 2190 turns off 1 5 4 Menus If there is no entry in the menu within three seconds the device returns to display mode 5 4 1 Select probe memory A memory 1 4 can be selected...

Page 22: ...nu key five times The reference value symbol appears with the set reference value such as C 58 Set the desired value with the plus or minus button Use the save button to save the set value The product...

Page 23: ...ly when calibrating the proximal probe Depending on the working direction of the probe the values can differ slightly for technical reasons however it is generally not necessary to recalibrate during...

Page 24: ...e the set value The product returns to display mode 5 4 5 Turning infrared data transmission on and off The infrared data transmission can be turned on and off ON OFF Press the menu key four times The...

Page 25: ...0 to 12 Normal prophylaxis such as fluoride toothpaste 13 to 24 Intensive prophylaxis such as fluoridation KaVo HealO zone 25 Minimally invasive restorative procedures Filling materials and intensive...

Page 26: ...Scan the teeth after cleaning but before fluoridation 2 Before scanning the teeth the teeth and interdental spaces should be dry since saliva can deflect the light especially in the interdental space...

Page 27: ...n the ring switch until you hear two beeps and the display appears 1 2 3 The device is individually adjusted to the patient The displayed value lies between 00 and 9 when the probe does not contact th...

Page 28: ...e last time the ring switch was press ed The tone sounds at a MOMENT display of 06 The higher the MOMENT value 06 99 the higher the frequency of the signal tone Note Foreign light sources can cause pr...

Page 29: ...max values F Briefly press the ring switch when the value of the investigated tooth is elevated to reset the PEAK value Scan the proximal area Note Dry the proximal area before use Hold the start but...

Page 30: ...ct concretions and calculus with the paro probe optional accessory The DIAGNOdent pen 2190 is designed to be used in connection with the Perio probe to support the user in diagnosing calculus or concr...

Page 31: ...he Paro probe into the PA pocket parallel to the tooth surface During the examination the Perio probe of the DIAGNOdent pen 2190 must contact the tooth surface and be slowly moved across it When the P...

Page 32: ...aque or calculus can adhere to the probe and influence the MOMENT values The MOMENT values do not change even though the probe is moved in and out of the pocket In this case the probe should be cleane...

Page 33: ...the process are required Likewise any deviation from the instructions by the person preparing the device should be carefully checked to see if it is effective and potential negative consequences shou...

Page 34: ...move major soiling directly after soiling with a single use paper towel Manual cleaning Clean all outer surfaces of the DIAGNOdent pen 2190 with a soft cloth and with one of the indicated disinfectant...

Page 35: ...sinfection Not applicable 6 1 4 Sterilization in a steam sterilizer in compliance with DIN EN 13060 sterilisation should directly follow cleaning Only the grip sleeve reference and probes can be steri...

Page 36: ...DIAGNOdent pen 2190 and direct the red laser beam on to a white piece of paper when the device is turned on Use the image on the paper to evaluate the probe Investigate the prismatic end of the proxi...

Page 37: ...opening Broken or scratched probe Replace the probe The given order sequence was not followed during the adjustment Repeat the adjustment Display shows ERR1 Checksum of program memory is incorrect Tur...

Page 38: ...approved by KaVo Presentation Material summary Mat No Light probe prox saphir mont 1 002 6970 Light probe fissure saphir mont 1 002 6967 Light probe Perio sapphire mont 1 002 8568 Gripping sleeve 1 0...

Page 39: ...tic interference The DIAGNODENT pen type 2190 is for use in an environment like the one cited be low The customer or user of the DIAGNOdent pen type 2190 should ensure that it is used in the correct e...

Page 40: ...2 These guidelines may not be applicable in every case The spread of electro magnetic waves is absorbed and reflected by buildings objects and people a The ISM frequency bands for industrial scientif...

Page 41: ...the minimum distance between portable and mobile HF telecommunications devices transmitters and the DIAGNOdent pen type 2190 depending on the output of the communication device as indicated below The...

Page 42: ...Instructions for use DIAGNOdent pen 2190 9 Information on electromagnetic compatibility 6 1 Setup methods according to DIN EN ISO 17664 42 44...

Page 43: ......

Page 44: ...1 003 6584 Fk 20141009 07 en...