Instructions for use DIAGNOdent pen 2190
9 Information on electromagnetic compatibility | 6.1 Setup methods according to DIN EN ISO 17664
9 Information on electromagnetic compatibility
Electromagnetic Transmissions
The DIAGNODENT pen type 2190 is for use in an environment like the one cited be‐
low. The customer or user of the DIAGNOdent pen type 2190 should ensure that it is
used in the correct environment.
Measurements of noise transmis‐
sions
Conformance
Electromagnetic environment -
guidelines
HF transmission according to
CISPR 11
Group 1
The DIAGNOdent pen type 2190
uses HF energy only for its internal
operation. Its HF transmission is
therefore very low, and it is improba‐
ble that neighbouring electronic de‐
vices will be disturbed.
Resistance to electromagnetic interference
The DIAGNODENT pen type 2190 is for use in an environment like the one cited be‐
low. The customer or user of the DIAGNOdent pen type 2190 should ensure that it is
used in the correct environment.
Interference immunity tests IEC 60601 test levels
Compliance level
Electromagnetic environ‐
ment - Guidelines
Electrostatic discharge
(ESD) according to IEC
61000-4-2
± 6 kV contact discharge
± 8 kV atmospheric dis‐
charge
± 2/4/6 kV contact dis‐
charge
± 2/4/8 kV atmospheric
discharge
Floors should be made of
wood or concrete or be fit‐
ted with ceramic tiles. If the
floor is fitted with synthetic
material, the relative hu‐
midity must be at least 30
%.
NOTE: V
T
is the alternating mains voltage before the test level is used.
Resistance to electromagnetic interference
The DIAGNODENT pen type 2190 is for use in an environment like the one cited be‐
low. The customer or user of the DIAGNOdent pen type 2190 should ensure that it is
used in the correct environment.
39 / 44