Instructions for use SMARTmatic S20 - 1.011.6750 SMARTmatic S20 K - 1.011.6751 SMARTmatic S20 S - 1.011.6752
7 Processing steps in accordance with ISO 17664 | 7.5 Drying
20 / 28
▶ For program settings as well as cleansers and disinfectants to be used, please re-
fer to the Instructions for Use of the washer disinfector.
▶ In order to prevent damage to the medical device due to residual fluid, make sure
that the inside and outside of the device is dry after the end of the cycle. Remove
any residual liquids from the interior and exterior of the medical device using com-
pressed air.
▶ Immediately after drying, lubricate the medical device with care agents from the
KaVo care system.
The drying procedure is usually part of the cleaning program of the washer disinfector.
Note
Please comply with the instructions for use of the washer disinfector.
7.5 Drying
Manual drying
▶ Clean the outside and inside with compressed air until no drops of water are visi-
ble.
Automated drying
The drying procedure is usually part of the cleaning program of the washer disinfector.
▶ Please comply with the instructions for use of the washer disinfector.
7.6 Care products and systems - Servicing
WARNING
Sharp bur in the medical device.
Risk of injury from sharp and/or pointed bur.
▶ Remove bur.
NOTICE
Improper service and care.
Premature wear and reduced product service life.
▶ Perform proper service regularly.
Note
KaVo only guarantees that its products will function properly when the care prod-
ucts listed as accessories are used, since these products have been tested for
proper use on our products.
7.6.1 Servicing with KaVo Spray
KaVo recommends servicing the product as part of the processing after each use, i.e.
after each cleaning, disinfection, and before each sterilization.
▶ Remove the insert.
▶ Cover the product with the Cleanpac bag.
▶ Plug the product onto the cannula, and press the spray button for one second.