Instructions for use SMARTmatic S20 - 1.011.6750 SMARTmatic S20 K - 1.011.6751 SMARTmatic S20 S - 1.011.6752
7 Processing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use
7 Processing steps in accordance with ISO 17664
7.1 Preparations at the site of use
WARNING
Hazard from incorrectly processed products.
Contaminated products are associated with an infection risk.
▶ Take suitable personal protective measures.
▶ Remove the tool from the medical device.
▶ Remove all residual cement, composite or blood immediately.
▶ Process the medical device as soon as possible after treatment.
▶ The medical device must be dry when transported to processing.
▶ Do not immerse in solutions or the like.
7.2 Non-fixing preliminary cleaning of the spray tube
Accessories required:
▪ Demineralized water 30 °C ± 2 °C (86 °F ± 3.6 °F)
▪ Nozzle needle
▪ Brush, e.g. medium-hard toothbrush
▪ Disposable syringe
Note
Before cleaning, take the hose off the spray clip and the contra-angle handpiece.
▶ Take the spray clip with spray tube off the dental instrument to improve the clean-
ing.
▶ Check the patency of the spray tube and clean it using the nozzle needle (Mat. no.
0.410.0931).
▶ Rinse the spray tube with at least 20 ml demineralized water using a disposable
syringe.
▶ If the spray tube is not patent after the manual rinsing procedure, the medical de-
vice must be replaced.
Brush the spray tube and hose under running tap water for at least 20 seconds using
a medium-hard toothbrush.
The non-fixing preliminary cleaning is a central constituent and must be carried out
prior to the automated processing.
In the cleaning and disinfecting device, validated internal cleaning of the spray tube
necessitates preliminary non-fixing cleaning.
17 / 28