Instructions for use SMARTmatic S20 - 1.011.6750 SMARTmatic S20 K - 1.011.6751 SMARTmatic S20 S - 1.011.6752
7 Processing steps in accordance with ISO 17664 | 7.4 Disinfection
▶ For program settings as well as cleansers and disinfectants to be used, please re-
fer to the Instructions for Use of the washer disinfector.
▶ In order to prevent damage to the medical device due to residual fluid, make sure
that the inside and outside of the device is dry after the end of the cycle. Remove
any residual liquids from the interior and exterior of the medical device using com-
pressed air.
▶ Immediately after drying, lubricate the medical device with care agents from the
KaVo care system.
The drying procedure is usually part of the cleaning program of the washer disinfector.
Note
Please comply with the instructions for use of the washer disinfector.
7.4 Disinfection
WARNING
Incomplete disinfection.
Infection hazard.
▶ Principally, KaVo recommends carrying out an final disinfection of the unpack-
aged item to guarantee the complete disinfection.
NOTICE
Using the disinfectant bath or chlorine-containing disinfectants.
Functional damage and property damage.
▶ Do not disinfect the device in the disinfection bath or with chlorine-containing dis-
infectants.
7.4.1 Manual external disinfection
KaVo recommends the following products based on material compatibility. The micro-
biological efficacy must be ensured by the disinfectant manufacturer and proven by
an expert opinion.
▪ CaviCide made by Metrex
Consumables required:
Cloths for wiping the medical device.
▶ Spray the disinfectant on a cloth, then wipe down the medical device and allow
the disinfectant to act in accordance with the instructions of the
disinfectant manufacturer.
▶ Comply with the instructions for use of the disinfectant.
7.4.2 Manual internal disinfection
Not applicable.
This product is suitable for automated disinfection only.
7.4.3 Automated external and internal disinfection
KaVo recommends washer disinfectors in compliance with ISO 15883-1, which are
operated using alkaline cleaning agents.
19 / 28