Instructions for use SMARTmatic S20 - 1.011.6750 SMARTmatic S20 K - 1.011.6751 SMARTmatic S20 S - 1.011.6752
7 Processing steps in accordance with ISO 17664 | 7.3 Cleaning
18 / 28
7.3 Cleaning
Note
Applicable only to the S20 S (using spray clips and hose) during automated external
and internal cleaning and disinfection:
Prior to automated processing, the spray clip must be connected to the dental in-
strument using the hose.
▶ Do not plug the spray clip onto the dental instrument.
NOTICE
Never process this medical device in an ultrasonic device.
Functional damage and property damage.
▶ Clean manually or in a washer disinfector only.
7.3.1 Manual external cleaning
Accessories required:
▪ Tap water 30
o
C ± 5
o
C (86
o
F ± 10
o
F)
▪ Brush, e.g. medium-hard toothbrush
▶ Brush under flowing tap water.
7.3.2 Automated external cleaning
KaVo recommends washer disinfectors in compliance with ISO 15883-1, which are
operated using alkaline cleaning agents.
▶ The program settings and cleansers and disinfectants that must be used can be
found in the instructions for use of the thermodisinfector.
▶ Directly after automated cleaning/disinfection, treat the medical device with the
care products and systems supplied by KaVo.
7.3.3 Manual internal cleaning
The interior of this product is not to be cleaned manually.
7.3.4 Automated internal cleaning
KaVo recommends washer disinfectors in compliance with ISO 15883-1, which are
operated using alkaline cleaning agents.