Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912
2 Safety | 2.1 Infection hazard
2 Safety
The instructions for use are a component of the product and must be read care-
fully prior to use and be accessible at all times.
The device may only be used in accordance with the intended use, any other
type of use is not permitted.
2.1 Infection hazard
Patients, users or third parties could be infected by contaminated medical
devices.
▶
Take suitable personal protective measures.
▶
Follow the instructions for use of the components.
▶
Before initial startup and after each use, reprocess the product and ac-
cessories appropriately.
▶
Carry out the reprocessing as described in the instructions for use. The pro-
cedure has been validated by the manufacturer.
▶
If you deviate from this procedure, it is essential to make sure that the re-
processing is effective.
▶
Reprocess the product and accessories appropriately before disposal.
2.2 Explosion hazard
Electrical sparks in the product can lead to explosion or fire.
▶
Do not use product in explosion hazard areas.
▶
Do not operate the product in an oxygen-enriched atmosphere.
▶
Do not use the product in the vicinity of flammable gases.
2.3 Electrical shock
Connecting a non-KaVo system to the medical device can lead to electrical
shock and injury to patient, user and third parties.
▶
During installation and operation of the product on treatment devices and
equipment from other manufacturers, comply with the provisions of "Pro-
tection from electrical shock," "Leakage current," and "Not grounding the
applied part" in accordance with IEC 60601-1.
▶
Make sure to combine the product only with a treatment centre / control
unit that has been released by KaVo.
▶
Make sure that the product combined with the dental treatment centre /
control unit meets the requirements in accordance with IEC 60601-1.
2.4 Technical condition
A damaged device or components could injure patients, users and third parties.
▶
Only operate devices or components if they are undamaged on the outside.
▶
Check that the device is working properly and is in satisfactory condition
before each use.
▶
Have parts with sites of breakage or surface changes checked by the Ser-
vice.