Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912
3 Product description | 3.2 Technical data
Motor electronics
Material nos. 0.641.1020,
0.641.1120, 0.641.1140,
0.696.0161 and 1.004.4317
Motor voltage rating
26 V DC
Motor speed rating
40,000 rpm at 26 V
Motor torque rating
2.6 Ncm
All handpieces and contra-angle handpieces can be mounted with the INTRA-
matic connection DIN EN ISO 3964.
▶
Inquire with the manufacturer for the connected loads on the device side
and a connection diagram for the motor side.
Operating mode
Exposure mode: 2 min. on, 5 min.
off
Air exit of the motor coupling
7.5 ± 1.5 NL/min.
Cooling air flow
32 ± 4 NL/min.
Cooling air
2.5 - 4.5 bar (36 – 65 psi)
Spray air
1.0 - 2.5 bar (15 – 36 psi)
Spray water
0.8 - 2.0 bar (12 – 29 psi)
Air requirements
Air quality in accordance with DIN
EN ISO 7494-2
dry, oil-free, dirt-free, non-contam-
inated
Air filter, supplied by customer
< 20 µm
Water requirements
Water quality in accordance with DIN
EN ISO 7494-2
Tap water
pH
7.2 - 7.8
Water filter, supplied by customer
< 80 µm
Note
Please comply with the Instructions for Use of the product to which the IN-
TRA LUX motor is to be attached. Check the suitability of the product for op-
eration with the INTRA LUX motor.
Note
Designed for operation on KaVo units:
▶
ESTETICA E80 / ESTETICA E70 / ESTETICA E50 / Status 1080 / Globus
1078 / Primus 1058
11 / 26
Summary of Contents for 1.000.6717
Page 1: ...Instructions for use INTRA K LUX 200 1 000 6717 INTRA K LUX 200 F 1 002 1912 ...
Page 24: ......
Page 25: ......