Instructions for use INTRA K-LUX 200 - 1.000.6717 | INTRA K-LUX 200 F - 1.002.1912
6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions at initial start-up
16 / 26
6 Checking for malfunctions and troubleshooting
6.1 Checking for malfunctions at initial start-up
CAUTION
Danger from hot lamp.
Risk of burn injury.
▶
Do not touch lamp after previous operation. Let the lamp cool off.
NOTICE
Missing or damaged O-rings.
Malfunction and premature failure.
▶
Make sure that all O-rings are on the coupling and are undamaged.
NOTICE
Inappropriate attachment of the handpiece to the coupling.
Damage to the O-rings.
▶
Take care when you attach the handpiece to the coupling.
Note
Stop working if the O-ring is missing or damaged.
The motor becomes too hot while idling:
▶
Check the amount of cooling air.
The motor runs too hot under load:
▶
Also check the handpiece.
When the speed drops or is uneven:
▶
Check the hose connection.
Motor coupling O-ring is missing:
▶
Replace O-rings.
6.2 Troubleshooting
6.2.1 Replacing the high-pressure lamp
CAUTION
Danger from hot lamp.
Risk of burn injury.
▶
Do not touch lamp after previous operation. Let the lamp cool off.
▶
Press the button
①
while pulling off the sleeve
②
in the direction of the ar-
row.
▶
Turn the ring
③
until the lamp is exposed
④
.
▶
Push the lamp
④
out of the holder in the direction of the arrow and lift it
out.