![Katronic Technologies KATflow 210 Operating Instructions Manual Download Page 28](http://html.mh-extra.com/html/katronic-technologies/katflow-210/katflow-210_operating-instructions-manual_657957028.webp)
KATflow 210
5 Commissioning
Dynamic
viscosity
0.000 ... 30000 mPa.s
Density
Only if user fluid selected
100 ... 2000 kg/m
3
Medium c-
speed
Only if user fluid selected
100 ... 3500 m/s
Temperature
-30 ... 300 °C
Liner Material
Select from list ↑↓
None, Epoxy, Rubber, PVDF, PP, Glass, Cement,
User (liner c-speed)
Liner thickness
Only if lining material selected
1.0 ... 99.0 mm
Liner c-speed
Only if lining material selected
600 ... 6553.0 m/s
Passes
Select from list ↑↓
Auto
1...16
Stored Setup
Load
Load from list
Save
Save to list – use alphanumeric keys to enter
name
Delete
Delete from list
Start
Measurement
Sensor type
Indication of sensor type and serial number if
automatically detected, otherwise
select from list ↑↓
As Setup Wizard
Sensor fre-
quency
SP1, only for special, unrecognised sensors
Wedge angle
SP2, only for special, unrecognised sensors
Wedge c-
speed 1
SP3, only for special, unrecognised sensors
Wedge c-
speed 2
SP4, only for special, unrecognised sensors
Crystal offset
SP5, only for special, unrecognised sensors
Spacing offset
SP6, only for special, unrecognised sensors
Zero flow offset
SP7, only for special, unrecognised sensors
Upstream off-
set
SP8, only for special, unrecognised sensors
K factor
Calibration factor (slope)
Sensor
placement
Adjust sensor position
Scope
Displays received acoustic pulse
Meas. Period
Change measurement period
(Note this will keep the setting in “Saver Mode”,
so will show minutes if “Yes”, seconds if “No”.)
Installation
Select channel
Channel 1, Channel 2
Pipe
Material
Select from pipe material list ↑↓
Outside dia-
meter
6 ... 6500 mm
28