![Katronic Technologies KATflow 210 Operating Instructions Manual Download Page 23](http://html.mh-extra.com/html/katronic-technologies/katflow-210/katflow-210_operating-instructions-manual_657957023.webp)
KATflow 210
4 Operation
Choose pipe material using cursor keys and
pressing <
ENTER
>.
Selecting “User” will allow manual entry of pipe
sound speed.
Enter outside pipe diameter using alphanumer-
ic keys and confirm by pressing <
ENTER
>.
Use key <
UP
> as character backspace clear
to correct for data entry errors.
If 0 is entered, an additional screen appears
that allows entering the pipe circumference.
Enter pipe wall thickness using alphanumeric
keys and confirm by pressing <
ENTER
>.
Use key <
UP
> as character backspace clear
to correct for data entry errors.
Enter inner pipe diameter using alphanumeric
keys and confirm by pressing <
ENTER
>.
The value that appears here will have been
calculted from the entered Outeside Diameter
(or Cicumference) and Wall thickness. Enter-
ing a new value here will recalculate the Out-
side Diameter.
Select fluid using cursor keys.
Confirm by pressing <
ENTER
>.
Selecting “User” will allow manual entry of
density, viscosity and medium sound speed.
Enter process temperature using alphanumeric
keys and confirm by pressing <
ENTER
>.
Use key <
UP
> as character backspace clear
to correct for data entry errors.
Select pipe lining material using cursor keys.
Confirm by pressing <
ENTER
>.
Selecting “User” will allow manual entry of liner
sound speed.
23