Forward Speed
0-100
Smaller value yields a lower maximum speed and vice versa
Reverse Speed
0-100
Smaller value yields a lower maximum speed and vice versa
Throttle Polarity
Yes/No
Yes = right speed lever is forward,
no= left speed lever is forward and vice versa
Automatic
Shutdown Time
0-20 Minutes
1-20 = minutes before shutdown time,
0 = no automatic shutdown
◎
Karma's Quality Control team sets the parameters of your central relay box (controller)
for optimum performance within the regulations of EN 12184. These parameters were
double checked by the QC team before your scooter left our factory.
◎
These parameters may be changed to suit certain individuals' specific needs and
requirements. If you need the parameters changed, please contact your dealer who
will put you in contact with our agent. They will know how to use the programmer
properly to "dial in" your scooter for your optimum usage. The programmer's effect
on the performance quality can been seen in the above chart.
WARNING
:
The safety limits that are set in EN 12184 tables 2 and 3 may be exceeded by changing
the parameters of the controller with a programmer. It is illegal to exceed these limits.
Your Karma agent and authorized service representative know these limits and will
adjust the performance quality within these limits to assure your safety, if you require the
factory set parameters to be altered.
KS-838 Static stability test result
Test Requirement
Min
Max
Static Stability Uphill
15
°
24
°
Static Stability Downhill
N/A
N/A
Static Stability Sideway
13
°
15
°
KS-848 Static stability test result
Test Requirement
Min
Max
Static Stability Uphill
17
°
27
°
Static Stability Downhill
N/A
30
°
Static Stability Sideway
18
°
21
°
※
Range (the maximum driving distance per complete charge) is estimated based on the following
conditions: 20
℃
(68
℉
), an 80kg (176lbs) driver, brand new fully charged batteries and a
constant driving speed of 10km/hr (6.2mph).