Overview of mitigating warnings
Instructions for use • TELE PACK + • CQH848_EN_V2.0_08-2022_IFU_CE-MDR
52
11 Overview of mitigating warnings
The original English warning text is as follows:
WARNING
To avoid the risk of electric shock, this equipment must only be connected to a sup-
ply mains with protective earth.
WARNING
Portable RF communications equipment (including peripherals such as antenna ca-
bles and external antennas) should be used no closer than 30 cm (12 inches) to any
part of the [ME EQUIPMENT or ME SYSTEM], including cables specified by the man-
ufacturer. Otherwise, degradation of the performance of this equipment could result.
WARNING
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary, this
equipment and the other equipment should be observed to verify that they are oper-
ating normally.
WARNING
Use of accessories, transducers and cables other than those specified or provided
by the manufacturer of this equipment could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in im-
proper operation.
WARNING
No modification of this equipment is allowed.