Application
Instructions for use • TELE PACK + • CQH848_EN_V2.0_08-2022_IFU_CE-MDR
36
Internally stored recordings can be removed from the internal storage only via USB export. If large
volumes of data are to be stored, a correspondingly large USB storage medium has to be available.
The current status of data transfer can be checked in the information area.
6.10 Patient data security
The system does not contain patient data as standard. The functions for managing patient data must
be activated first, e.g., input, access, display.
Upon activating the functions, the user assumes responsibility for all patient-related data that
are managed in accordance with the applicable national data protection laws.
In the Patient Data Privacy settings, all settings concerning patient information can be configured.
Patient information is available only if the corresponding settings have been made. The configuration
of the Patient Data Privacy menu is password-protected by default.
The Patient Data Privacy menu is divided into 3 areas:
–
Patient handling
–
Labeling
–
Export mode
6.10.1 Patient handling
Worklist setup
The worklist setup specifies whether patient data is allowed to be used and how the patient data can
be processed on the TELE PACK+, whether this is via manual entry, patient worklist, or network
worklist providers.
Select the worklist setup:
–
No Worklist / Patient Entry (selected by default)
–
Local Worklist
–
Worklist Provider
When working without entering patient data, the data in the information panel must be exported via
Transfer to USB.
Capture tools
Whether and how images and videos are to be recorded is defined in the capture tools:
–
No Capture
–
Capture Allowed (selected by default)
If Capture Allowed is selected, the following settings are available:
–
Recording of images and videos with patient information
–
Anonymous recording
Storage Limit
The storage limit specifies the period during which data sets are stored on the device. Patient data
sets in the worklist and media files are deleted if this is configured. The data cannot be recovered.
Select storage limit:
–
Infinite
–
Power cycle
–
1 day (selected by default)
–
7 days