Cleaning, Disinfection and
Sterilization Instructions (Continued)
Note:
If immediate cleaning is not feasible,
place the camera head in a container and
soak for no longer than 30 minutes in a
diluted neutral/mild pH enzymatic cleaning
solution (e.g. Enzol).
Manual Cleaning
Recommended Water Quality:
Utility Water
Caution:
WEAR PROTECTIVE GLOVES,
CLOTHING, AND A FACE MASK FOR
CLEANING OF CONTAMINATED INSTRU-
MENTS.
1.
Thoroughly rinse the camera head
using a commercial overhead sprayer
nozzle with cold water to remove all
gross debris. Rinse the camera head
unit for a minimum of 1.5 minutes, pay-
ing special attention to the knobs,
crevices, and seams. Then rinse the
connecting cable for a minimum of 30
seconds.
2.
Completely immerse the camera head
in a diluted mild/neutral pH enzymatic
cleaning solution (e.g. Enzol). Keep
immersed for a minimum of 1 minute.
3.
While immersed, use a soft bristle
brush (P/N: 27652/3) to brush all the
surfaces of the camera head unit. Pay
close attention to the seams between
the instrument coupler, focusing ring
and zoom ring. Do not brush the glass
window. Brush for a minimum of 1
minute.
4.
Then, use a low-lint wipe to wipe the
entire length of the connecting cable
and the glass window of the camera
head.