3-6
05.13 / V1.0
UNIDRIVE
®
S III
3.4.2. Error message description
Error code
Error message
Description
1
E01
Memory error
2
E02
Motor board defective
3
E03
Hardware configuration error
4
E04
Software configuration error
5
E05
---
6
E06
---
7
E07
---
8
E08
---
9
E09
Motor M1 defective
10
E10
Motor M2 defective
11
E11
No footswitch
12
E12
No motor
13
E13
Motor M1 overheated
14
E14
---
15
E15
Motor M2 overheated
16
E16
Irrigation pump not closed
17
E17
Motor M1 not supported
18
E18
Motor M2 not supported
19
E19
Motor M1 stopped (blocked)
20
E20
Motor M2 stopped (blocked)
21
E21
No SCB
3
WARNING: Always unplug the equipment before performing any repairs.
3
WARNING: All devices must be disinfected prior to repair.
2
CAUTION:
Repairs may only be performed by qualified technicians trained in electrical or
electronic engineering, in compliance with the relevant occupational, safety and accident
prevention regulations.
Safety Testing based on IEC EN 62353, IEC / UL 60601-1, whichever may apply, must be
performed after servicing has been completed.
Summary of Contents for 20 7010 20-1
Page 4: ......
Page 8: ......
Page 9: ...I UNIDRIVE S III ENT UNIDRIVE S III ENT UNIDRIVE S III ENT 40701620 1 40701620 1 40701620 1 ...
Page 10: ......
Page 12: ......
Page 127: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 134: ......
Page 146: ......
Page 148: ......
Page 158: ......
Page 166: ......
Page 168: ......
Page 169: ......