101
Table 202
Guidance and manufacturer’s declaration – electromagnetic immunity
The UNIDRIVE
®
S III ENT model
40
70
16
20-1 is intended for use in the electromagnetic environment specified below.
The customer or user of the UNIDRIVE
®
S III ENT model
40
70
16
20-1 should ensure that it is used in such an environment.
Immunity test
EN/IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic Discharge (ESD)
acc. to IEC 61000-4-2
± 6 kV contact
± 8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
acc. to IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
2 kV/1 kV
Mains power quality should be that of a typical commercial
or hospital environment.
Surge
acc. to IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
2 kV/1 kV
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
acc. to IEC 61000-4-11
<5% U
T
*
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
0 V, 10 msec
92 V, 100 msec
168 V, 500 msec
0 V, 5 sec
Mains power quality should be that of a typical commercial
or hospital environment.
If the user of the equipment or system requires continued
operation during power mains interruptions, it is recom-
mended that the equipment or system be powered from
an uninterruptible power supply.
Power frequency
(50/60 Hz)
magnetic field
acc. to IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels char-
acteristic of a typical location in a typical commercial or
hospital environment.
Anexo
Appendix
Anhang
Summary of Contents for 20 7010 20-1
Page 4: ......
Page 8: ......
Page 9: ...I UNIDRIVE S III ENT UNIDRIVE S III ENT UNIDRIVE S III ENT 40701620 1 40701620 1 40701620 1 ...
Page 10: ......
Page 12: ......
Page 127: ......
Page 129: ......
Page 130: ......
Page 131: ......
Page 134: ......
Page 146: ......
Page 148: ......
Page 158: ......
Page 166: ......
Page 168: ......
Page 169: ......