24
4. Specifications
Product name Air Flow Transducer
Model
6333-0G
Air velocity
measurement
range
0.01 to 25 m/s (For connected probe: 0975-00, 0975-21, 0975-31)
0.01 to 30 m/s
(For connected probe: 0976-03, 0976-04, 0976-05, 0976-07, 0976-13,
0976-14, 0976-15, 0976-17)
0.01 to 50 m/s
(For connected probe: 0972-00, 0973-00, 0975-09, 0975-10)
Temperature
measurement
range
0 to 60℃ (For connected probe: 0975-31)
0 to 100℃ (For connected probe: 0975-21)
0 to 120℃ (For connected probe: 0975-05)
Humidity
measurement
range
5 to 95% (For connected probe: 0975-31)
Resolution
Air velocity: 0.01 m/s, Temperature: 0.1℃, Humidity: 0.1%RH
Unit
m/s・FPM (Factory setting: m/s), ℃・°F (Factory setting: ℃), %RH
The DIP switch enables to change between m/s and FPM, and ℃ and °F.
Analog output
The output conversion DIP switch enables to change between voltage
output and current output.
Factory setting is current output.
Current output: DC 4 to 20 mA (Maximum load: 250 Ω)
Voltage output: DC 0 to 5 V
Interface
RS485
Power source
Power source is selectable from the AC adapter and external power source.
AC adapter: 12V/2A, External power source (Terminal base): 12V/2A to
24V/1A
Operating
environment
Temperature
0 to 60℃ (non-condensing)
Humidity
2.0%RH to 98%RH
Storage
environment
Temperature
-10 to 60℃ (non-condensing)
Humidity
2%RH to 95%RH
Dimensions
Approximately W100.0 × D60.0 × H25.0 mm
Weight
Approximately 200 g
Measurement
software
Air Flow Transducer Measurement Software (optionally available)
Standard
accessories
User’s manual: 1 copy, Terminal block: 1 EA
Warranty
period
One year from date of purchase
Summary of Contents for 6333
Page 1: ...i Air Flow Transducer Model 6333 User s Manual ...
Page 2: ......
Page 11: ...9 1 2 External Dimension Unit mm The dividing terminal block is excluded ...
Page 30: ...28 ...
Page 32: ...30 ...
Page 37: ...35 ...
Page 40: ...38 ...