<Zero Adjustment Procedure>
DISPLAY
PROCEDURE
Press
key in the Measurement Mode to select pressure
measurement.
Each time you press the
key, the measurement item changes
in the order of Air Velocity -> Air Temperature -> Pressure.
If you hold down
key for over 2 seconds, you will hear a
long beep sound, and the pressure will be indicated as “0.00”.
<Connecting Pressure Tube>
Connect the pressure tube to the (+) or (-) port.
Connect the other end of the tube to the port (e.g. duct) where the pressure is to be measured.
When the pressure to be measured is positive, connect the tube to the positive (+) port. And when it is negative,
connect the tube to the negative (-) port.
2. Getting Started
10
In order to make an accurate measurement, make
sure that the tube does not have any holes, which
could cause air leakage. Also if the tube is bent, the
measurement will not be accurate.
kPa
kPa
Positive (+) Port
Negative (-) Port
Pressure Port