2 Safety Instructions
Please study this manual thoroughly and you will find the Kanmed BabyWarmer BW3 to be both safe and
easy to use.
Intended use
The intended use of the Kanmed BabyWarmer BW3 is to keep newborn and premature babies normo-
thermic.
Kanmed BabyWarmer BW3 is designed for hospital use only.
It must be used according to the instructions in the manual and according to established clinical routines
and by qualified personnel.
Essential Performance
The automatic system supervision will safeguard any malfunction that may have influence on the performance of the
system. If used according to this instruction manual, and if the warning information is considered, no unacceptable
patient risks should occur. However, always monitor the patient's vital signs and adapt the use after the patient's actual
requirements.
Warning
•
Proper Use.
To ensure optimal performance and to prevent incorrect use, the user manual must be
read completely before the system is being used. Only use genuine Kanmed BabyWarmer parts
together with the BW3 control unit.
•
System Malfunction.
If the BabyWarmer BW3 self test does not perform correctly or the system
displays an alarm or error message, or the Control Unit has been dropped, received mechanical
damage etc., it must be examined by a qualified technician before being used. If it is suspected that
the BabyWarmer BW3 is not operating properly, immediately consult a qualified technician for advice
and do not use it.
•
Never use the BabyWarmer BW3 without a Kanmed Water or Gel Mattress!
•
Electrical Hazard
The system must always be connected to a mains power outlet with proper
protective grounding. Always remove the mains power cable before cleaning the Control Unit.
•
Body temperature.
The temperature monitor is intended for added safety purposes only, please use
a dedicated hospital grade precision thermometer, and a method in accordance to hospital procedures
for precise measurements of the
patient’s
actual body temperature.
•
Never place a baby face down on the Kanmed Mattresses or in the BabyWarmer Nest unless this is
ordered by responsible staff.
•
A cold mattress, or a mattress which is cooling down due to the heating being switched off, will
decrease the body temperature of the baby. Likewise, a too warm mattress can induce fever
.
•
Trans dermal medications (patches) can increase drug delivery, resulting in possible harm to patient
when used with warming devices.
Caution
•
Check the Water Mattress daily for leaks.
•
Be aware that the acoustic alarm has a reduced sound level of 55dBA (adapted for use in silent
environments, e.g. a neonatal ward).
•
Always fill the Water Mattress to the required level before the unit is brought into use. Preferably fill
with lukewarm water at around 35
0
C.
•
Replace the Water Mattress periodically as indicated on the Water Mattress (after about 12 months of
use).
•
Don't fold or bend the Heating Pad sharply, don't pull the Heating Pad cable or use it for carrying the
Heating Pad.
•
Make sure that the text side of the Heating Pad is facing the Mattress.
•
Make sure that there are at least 2 drain holes in the bottom of the bed if using the Water Mattress.
•
Make sure that the adjusting string is placed outside the Baby Nest to prevent it from interfering with
the baby.
•
Check the temperature of the baby regularly. Check the mattress temperature regularly.
•
Continue to use appropriate monitoring of vital signs.
•
Before first use, please clean everything according to your standard hospital routines and to
instructions in the user manual. Always wash a new re-usable Baby Nest before first use.
•
Please note that the control unit BW3 shall be placed standing on a flat surface, or mounted with the
front end up, having the display clearly visible.
•
The BW3 may not be interfaced with any other electrical system. If this is done a new “system” is
created per definition, and the safety classification of the BW3 may be effected. Please be aware that
use of HF surgical instruments or similar devices may interfere with the BW3 and may require special
precautions regarding potential equalization etc.
•
The Kanmed Gel Mattress may not be used without its integrated aluminium-plate.