KANE450
manual
Page
7
Backlight
Press and hold
to toggle the display's backlight
Note: use of the backlight increase the current drain on the batteries
Tasklight
Comes on in conjunction with the display's backlight... see above.
Switching
PUMP on / off
The analyser normally operates with the pump on. Press PUMP
quickly
to switch the pump off and on.
When the pump is switched off, the analyser displays "PUMP OFF"
approx every 30 seconds.
NOTE: the pump will not switch off if the CO reading is above
20ppm .
This helps to protect the CO sensor from damage.
NOTE: the pump cannot be switched off with the rotor set to
-----------MENU.
Zeroing the
pressure sensor
Press and hold the PUMP key until the top line display shows ZERO
CAL.
“Freezing” the
display
Press and hold the HOLD / ENTER key until a second beep is heard to
freeze all readings. The display flashes and readings can be printed by
pressing PRINT key. Press and hold the HOLD / ENTER key again until
a second beep is heard to go back to “live” measurements.
The HOLD function is inhibited when the display LEDs are flashing
Storing a set of
readings
Press and hold the HOLD/ENTER key for approx. 4 seconds.
Note: this STORE function is inhibited if the display LEDs are flashing
Note: this STORE function is inhibited in normal operation if the pump is
switched off.
Changing the
display's active
line
Press and release the HOLD/ENTER key until the illuminated LEDs point
at the display line that you wish to change. While the LEDs are flashing
turn the rotary dial to display the parameter that you want. If the rotor is
not turned the active line will go back to its previous setting after approx.
5 secs.