7. MEASURING 

MODES 

 
MEASURING FLUE GASES 

 
After the countdown is finished and the analyser is correctly set up, put its’ flue probe into the 
appliance’s sampling point. The ideal sampling point is at least two flue diameters 
downstream of any bend.  
 
Put the probe tip in the flue center. With balanced flues, make sure the probe is positioned far 
enough into the flue so no air can ‘back flush’ into the probe. 
 
 

 

 
 
Use the probe’s depth stop cone to fix it in flue diameters from 

1

/

4

 to 

4

/

5

 inch, 6 to 21 mm. Its 

maximum operating temperature is 1112

°

F / 600

°

C. 

 
Make sure you do not exceed the analyser’s operating specifications. In particular: 
 

 

Do not exceed the flue probe’s maximum temperature 

 

Do not exceed the analyser’s internal temperature operating range  

 

Do not put the analyser on a hot surface 

 

Do not exceed the water trap’s levels 

 

Do not let the analyser’s particle filter become dirty and blocked 

 
View the displayed data to ensure that stable operating conditions have been achieved and the 
readings are within the expected range. 
 
Press PRINT  or  press HOLD first to “freeze” or store the readings before printing 

Summary of Contents for 450

Page 1: ...7 KANE450 Flue Gas Analyser Kane International Ltd Kane House Swallowfield Welwyn Garden City Hertfordshire AL7 1JG Tel 44 0 1707 375550 Fax 44 0 1707 393277 E mail sales kane co uk www kane co uk Sto...

Page 2: ...FROM MENU 9 6 MAIN MENU FUNCTIONS 12 7 MEASURING MODES 16 22 MEASURING FLUE GASES 16 STANDARD COMBUSTION PRINTOUT 17 STORING A DRAUGHT MEASUREMENT TO PRINT WITH A COMBUSTION TEST 18 MEASURING PRESSUR...

Page 3: ...ective rubber sleeve with a magnet for hands free operation and their flue probes work on all combustion appliances and take ambient air measurements The large display shows 4 readings at a time and a...

Page 4: ...are recommended WARNING The battery charger unit must only be used when NMiH batteries are fitted NMiH batteries are entirely suitable for top up charging at any time even for shorts periods Alkaline...

Page 5: ...cts combustion gases that may be toxic in relatively low concentrations These gases are exhausted from the back of the instrument This analyser must only be used in well ventilated locations by traine...

Page 6: ...plug to the analyser during countdown the measured temperature from the flue probe will be used as the inlet temperature If neither probe is connected during countdown the analyser s internal ambient...

Page 7: ...lay shows ZERO CAL Freezing the display Press and hold the HOLD ENTER key until a second beep is heard to freeze all readings The display flashes and readings can be printed by pressing PRINT key Pres...

Page 8: ...display pressure meter thermometer with the pump off and inhibited The display will show pressure flue temp inlet temp differential temp The rotor display indications will now be locked apart from MEN...

Page 9: ...low a pre set limit the display flashes LOW BAT every 10 seconds See above to change the batteries or re charge if applicable TIME Displays Time LOSS Displays calculated losses when O2 values are less...

Page 10: ...aft value in user selected units X AIR Displays calculated excess air when O2 values are less than 18 EFF Displays calculated efficiency when O2 values are less than 18 EFFn EFFg or EFFc as selected b...

Page 11: ...the change and exit If you rotate the dial before pressing ENTER the instrument will beep continuously You must rotate the dial back to MENU and press ENTER before you can continue In STORE View mode...

Page 12: ...ficiency calculation time and date setting temperature unit gas units oxygen referencing language exit PRESSURE Smoothing resolve number of decimal points resolution number of decimal points resolutio...

Page 13: ...tt Condensing Gross efficiency calculation SET TIME xx yy zz HH MM SS set time values SET DATE DD MM YY MM DD YY YY MM DD choose date format then set date values C F C F centigrade fahrenheit PPM MG p...

Page 14: ...TEST xxx enter test number to be viewed then use rotor to select display If less than 9 set of readings are stored rotate the dial then use the up down buttons to change the LOG number AUTO STO YES N...

Page 15: ...620 6677 KANE450 manual Page 15 SCREEN CONTRAST SET xx lighten or darken the display AUX Selected Fuel Pressure units COM Date CO air free O2 Reference value assign the AUX rotor position HEADER Heade...

Page 16: ...s depth stop cone to fix it in flue diameters from 1 4 to 4 5 inch 6 to 21 mm Its maximum operating temperature is 1112 F 600 C Make sure you do not exceed the analyser s operating specifications In...

Page 17: ...3456789 Kane s serial number sales kane co uk user header 1 KANE INTERNATIONAL user header 2 DATE 12 12 05 TIME 09 53 23 FUEL NATU GAS O2 5 1 CO2 9 0 CO ppm 400 NO ppm n f NO sensor not fitted FLUE o...

Page 18: ...lues will be logged As this occurs the top line display briefly shows DRAUGHT Whilst the instrument is in this mode activating LOG again will overwrite the stored reading The stored reading is cleared...

Page 19: ...he test the analyser waits 60 secs to allow temperature stabilisation then records and displays the pressure reading at the start of a 120 second countdown The live pressure reading is displayed durin...

Page 20: ...ial thermometer ideal for measuring flow and return temperatures or for setting up hot water temperature flow on a combi boiler Standard Temperature Report this mode is accessed via the MENU function...

Page 21: ...conducted where the analyser automatically records 16 sets of 1 minute samples and prints the results as below Standard Room CO Report this mode is accessed via the MENU function K450 1 0 Serial No 12...

Page 22: ...ares co Tel 0161 620 6677 KANE450 manual Page 22 MEASURING FLUE DRAUGHT STANDARD DRAUGHT REPORT K450 1 0 Serial No 123456789 sales kane co uk KANE INTERNATIONAL Draught TIME 15 58 12 12 05 PRS mBar 0...

Page 23: ...robe from the flue THE PROBE WILL BE HOT and let it cool Do not put the probe in water which will be sucked into the analyser damaging its pump and sensors When the analyser s readings return to ambie...

Page 24: ...not giving correct output Fuse blown in charger plug Analyser does not respond to flue gas Particle filter blocked Probe or tubing blocked Pump not working or damaged with contaminants Net temperatur...

Page 25: ...damage Local regulations may require more frequent re calibration In the UK Kane International has service facilities at Atherton near Manchester Tel 01942 873434 the primary service centre for UK cu...

Page 26: ...1112 F Pressure Measurement 0 1 Pa 0 1 Pa 1 Pa 0 01 hPa 0 5 Pa 3 Pa 3 of reading 3 of reading 20 Pa 100 Pa 2000 Pa 80 hPa Gas Measurement Oxygen Carbon Monoxide KANE450CO KANE450COH Nitric Oxide NO C...

Page 27: ...ss steel shaft type K thermocouple and 3m 6ft long neoprene hose Ambient Operating Range 0 C to 40 C 32 104 F 10 to 90 RH non condensing Storage Capacity LOG REPORT 255 sets of test results in volatil...

Page 28: ...ck all readings are as expected A level of disturbance is acceptable If not acceptable adjust the analyser s position to minimize interference or switch off if possible the offending equipment during...

Page 29: ...as not set to zero correctly switch off instrument and try again C0m Carbon Monoxide displayed in mg m3 COn Carbon Monoxide in ppm normalised to the O2R value CO Carbon Monoxide in mg m3 normalised to...

Page 30: ...n each combustion test is printed Note When changing the batteries on the instrument the memory will store the date and time for up to one minute if outside this time it may be necessary to re enter t...